P385 – Exit Slit Unit B07 DLS

Application

Exit slit unit for soft X-ray synchrotron radiation

Year of delivery

2014

Installation site

Beamline B07 at Diamond Light Source, Didcot, United Kingdom

Design Features

  • Double slit mechanism – vertically and horizontally defining slits.
  • Motorized and encoded slits opening, symmetric opening, non-linear opening characteristics.

  • OFHC slit blades.
  • Electrical isolation of all slit blades.
  • Motorized and encoded slit unit translation along beam axis.
  • Bias plate with adjustable YAG screen.

  • Adjustable support structure.

Performance Features

  • Precision and slit opening repeatability smaller than 100 nm.
  • Parallelism smaller than 0.9 µm over 10 mm edge length.
  • Straightness smaller than 0.5 µm over 10 mm edgle length.
  • Very small deviation (submicrometer) of the slit opening calibration curve from the theoretical non-linear slit opening function.

Outer Dimensions

P385 - Outer dimensions - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Main flanges

DN160CF cluster flanges

Material

1.4301

Base pressure

10-10 mbar range

Translation along beam axis

Range

± 100 mm

Resolution (design value)

0.015 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Support structure
adjustment ranges

Height

± 10 mm

Lateral

± 10 mm

Longitudinal

± 10 mm

Horizontally defining slit (HDS)

Slit blades

Material

OFHC

Straightness

< 0.25 µm over 10 mm (1σ)

Parallelism

0.03 µm /mm
(15 µm slit opening)
0.03 µm /mm
(25 µm slit opening)
0.05 µm /mm
(50 µm slit opening)

Slit opening

Range

6.7 µm to 10.4 mm

Repeatability

± 0.04 µm
(up to 350 µm slit opening)

Resolution (design value)

0.006 µm /motor fine step
(1/16, recommended)
0.05 µm /encoder count

Vertically defining slit (VDS)

Slit blades

Material

OFHC

Straightness

< 0.15 µm over 10 mm (1σ)

Parallelism

0.06 µm /mm
(15 µm slit opening)
0.06 µm /mm
(25 µm slit opening)
0.08 µm /mm
(50 µm slit opening)

Slit opening

Range

4.4 µm to 11 mm

Repeatability

± 0.03 µm
(up to 350 µm slit opening)

Resolution (design value)

0.006 µm /motor fine step
(1/16, recommended)
0.05 µm /encoder count

Performance Test Results

Slit width calibration (HDS)
P385 - Slit width calibration (HDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit opening repeatability (HDS)
P385 - Slit opening repeatability (HDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades parallelism (HDS)
P385 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades parallelism (HDS)
P385 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades parallelism (HDS)
P385 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit width calibration (VDS)
P385 - Slit width calibration (VDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit opening repeatability (VDS)
P385 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades parallelism (VDS)
P385 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades parallelism (VDS)
P385 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Slit blades parallelism (VDS)
P385 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH
Residual gas analysis
P385 - Residual gas analysis - Exit slit unit for soft X-ray synchrotron radiation at Beamline B07 at Diamond Light Source, Didcot, United Kingdom - Bestec GmbH