P456 – 4″ HV sputter deposition HZDR

Application

HV sputter deposition system for thin film and multilayer deposition at 4″ substrates

Year of delivery

2017

Installation site

HZDR, Dresden-Rossendorf, Germany

Design Features

  • HV magnetron sputter deposition system in face to face sputter down configuration.
  • Up to six 4″ magnetrons in face to face configuration with possibility of low pressure and low power sputtering.
  • Motorized sample manipulator with maximal sample temperature well above 900°C.
  • Integrated bake out system.
  • Load lock chamber with storage.

Special Features

  • System is prepared to be added to a cluster tool via second transfer port at sputtering chamber.
  • To the load lock chamber a second process chamber can be added.
  • Different sample sizes from 4″ wafer down to 10mm x 10mm samples can be handled (using different kind of sample adapters).

Outer Dimensions

Technical specifications and performance values

General

Sputtering chamber

Size

800 mm diameter, about 500 mm height

Material

stainless steel

Load lock chamber

Size

200 mm diameter, about 500 mm height

Material

stainless steel

Vacuum

Sputtering chamber

Base pressure

< 5 *10-8 mbar

Pump down time

6 minutes to < 10-5 mbar

Chamber pumping

Turbo pumping stage, chamber lid differentially pumped by dry foreline pump

Bake out

< 120°C

Load lock chamber

Base pressure

< 7 * 10-8 mbar

Pump down time

4 minutes to < 10-5 mbar

Chamber pumping

Turbo pumping stage with dry foreline pump

Manipulator features

Sputtering chamber

Sample size

diameter max. 4″ substrate

Motion axes

motorized manipulator arm rotation (continous rotation and positioning)

Temperatures

Room temperature (not stabilized) up to 1000°C at sample (short time heating) / > 850°C at sample (long time heating)

Load lock chamber

(Sample storage)

Storage size

10 sample holders

Sample size

diameter max. 4″ substrate

Motion axes

2 manual axes (rotation, z tranlsation)

Rotation axis equipped with an air side idexer plate for easy and fast sample loading via access door or transfer rod

Performance test results

Chamber pump down
Long time sample heating